左心室四极导线在心脏再同步化治疗中的临床应用Clinical application of left ventricular quadripolar lead in cardiac resynchronization therapy
邹彤,佟佳宾,施海峰,陈浩,刘俊鹏,张倩,王志蕾,金鑫,吴素娟,曲祎玫,杨杰孚
摘要(Abstract):
目的初步探讨左心室四极导线在心脏再同步化治疗(cardiac resynchronization therapy,CRT)中的临床应用。方法选取北京医院心内科2014年1月至2015年9月连续收治的CRT新植入患者共40例,分为四极导线组(13例)和双极导线组(27例)。比较两组患者的手术时间、X线曝光时间、心功能改善及并发症发生情况等的差异。结果两组患者左心室导线置入时间、手术时间、X线曝光时间、左心室导线位置比较,差异均无统计学意义(均P>0.05)。术后6个月,四极导线组左心室射血分数(LVEF)[(38.2±4.1)%比(33.3±5.0)%,P=0.026]、纽约心脏病协会(NYHA)心功能分级[(2.4±0.4)级比(2.8±0.6)级,P=0.037]、CRT有反应率[11例(84.6%)比18例(66.7%),P=0.017]和心功能改善≥1级[10例(76.9%)比18例(66.7%),P=0.019]均优于双极导线组。此外,对于QRS波时限的缩短[(141.7±14.0)ms比(154.4±15.3)ms,P=0.006],四极导线组优势更为显著;而对于左心室舒张末期内径改善的比较,两组间差异无统计学意义(P>0.05)。四极导线组在围术期对于膈神经刺激和阈值的管理优于双极导线组。结论短期随访提示左心室四极导线在血流动力学改善、CRT有反应率和临床疗效方面均优于双极导线。
关键词(KeyWords): 心脏再同步化治疗;左心室四极导线;膈神经刺激;NYHA心功能分级
基金项目(Foundation): 国家自然科学基金项目(81170171)
作者(Author): 邹彤,佟佳宾,施海峰,陈浩,刘俊鹏,张倩,王志蕾,金鑫,吴素娟,曲祎玫,杨杰孚
参考文献(References):
- [1]Cleland JG,Daubert JC,Erdmann E,et al.The effect of cardiac resynchronization on morbidity and mortality in heart failure.N Engl J Med,2005,352(15):1539-1549.
- [2]Bristow MR,Saxon LA,Boehmer J,et al.Cardiacresynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure.N Engl J Med,2004,350(21):2140-2150.
- [3]Moss AJ,Hall WJ,Cannom DS,et al.Cardiac-resynchronization therapy for the prevention of heart-failure events.N Engl J Med,2009,361(14):1329-1338.
- [4]Tang AS,Wells GA,Talajic M,et al.Cardiac-resynchronization therapy for mild-to-moderate heart failure.N Engl J Med,2010,363(25):2385-2395.
- [5]van Bommel RJ,Bax JJ,Abraham WT,et al.Characteristics of heart failure patients associated with good and poor response to cardiac resynchronization therapy:a PROSPECT(Predictors of Response to CRT)sub-analysis.Eur Heart J,2009,30(20):2470-2477.
- [6]Biffi M,Moschini C,Bertini M,et al.Phrenic stimulation:a challenge for cardiac resynchronization therapy.Circ Arrhythm Electrophysiol,2009,2(4):402-410.
- [7]Champagne J,Healey JS,Krahn AD,et al.The effect of electronic repositioning on left ventricular pacing and phrenic nerve stimulation.Europace,2011,13(3):409-415.
- [8]Jastrzebski M,Bacior B,Wojciechowska W,et al.Left ventricular lead implantation at a phrenic stimulation site is safe and effective.Europace,2011,13(4):520-525.
- [9]张澍,黄德嘉,华伟,等.心脏再同步治疗慢性心力衰竭的建议(2013年修订版).中华心律失常学杂志,2013,17(4):247-261.
- [10]Singh JP,Klein HU,Huang DT,et al.Left ventricular lead position and clinical outcome in the multicenter automatic defibrillator implantation trial-cardiac resynchronization therapy(MADIT-CRT)trial.Circulation,2011,123(11):1159-1166.
- [11]Osca J,Alonso P,Cano,et al.The use of quadripolar left ventricular leads improves the hemodynamic response to cardiac resynchronization therapy.Pacing Clin Electrophysiol,2015,38(3):326-333.
- [12]Thibault B,Dubuc M,Khairy P,et al.Acute haemodynamic comparison of multisite and biventricular pacing with a quadripolar left ventricular lead.Europace,2013,15(7):984-991.
- [13]曹芳英,柳景华.左心室多位点起搏心脏再同步化治疗研究进展.中国介入心脏病学杂志,2016,24(2):110-113.
- [14]Cleland J,Freemantle N,Ghio S,et al.Predicting the long-term effects of cardiac resynchronization therapy on mortality from baseline variables and the early response a report from the CAREHF(Cardiac Resynchronization in Heart Failure)Trial.J Am Coll Cardiol,2008,52(6):438-445.
- [15]Jessup M,Abraham WT,Casey DE,et al.2009 focused update:ACCF/AHA Guidelines for the Diagnosis and Management of Heart Failure in Adults:a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines:developed in collaboration with the International Society for Heart and Lung Transplantation.Circulation,2009,119(14):1977-2016.
- [16]Stevenson WG,Hernandez AF,Carson PE,et al.Indications for cardiac resynchronization therapy:2011 update from the Heart Failure Society of America Guideline Committee.J Card Fail,2012,18(2):94-106.
- [17]Mc Murray JJ,Adamopoulos S,Anker SD,et al.ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure2012:The Task Force for the Diagnosis and Treatment of Acute and Chronic Heart Failure 2012 of the European Society of Cardiology.Developed in collaboration with the Heart Failure Association(HFA)of the ESC.Eur Heart J,2012,33(14):1787-1847.
- [18]De Pooter J,El Haddad M,Timmers L,et al.Different Methods to Measure QRS Duration in CRT Patients:Impact on the Predictive Value of QRS Duration Parameters Ann Noninvasive Electrocardiol,2016,21(3):305-315.
- [19]Boriani G,Connors S,Kalarus Z,et al.Cardiac resynchronization therapy with a quadripolar electrode lead decreases complications at 6 months:results of the MORE-CRT randomized trial.Jacc Clinical Electrophysiology,2016,2(2):212-220.