无载体西罗莫司洗脱支架在小型猪冠状动脉模型中预防冠状动脉支架再狭窄的实验研究Evaluation of a novel polymer-free sirolimus-eluting stent in a porcine model
陈明,郑博,彭红玉,王新刚,霍勇
摘要(Abstract):
目的评价新型无载体西罗莫司药物洗脱支架(polymer-free sirolimus-eluting stent,PFSES)在小型猪冠状动脉模型中抑制新生内膜增殖的有效性和安全性。方法金属裸支架(baremental stent,BMS,n=13)、无载体纳米微孔裸支架(polymer-free bare mental stent,PFBMS,n=13)、聚合物载体西罗莫司洗脱支架(polymeribased sirolimus-eluting stent,PSES,n=13)以及PFSES(n=13)被分别随机置入26头小型猪的前降支(n=26)和回旋支(n=26)。支架置入28天和90天后,复查冠状动脉造影评价管腔丢失。90天后处死部分动物(n=12)行塑料包埋硬组织切片染色组织形态学分析。结果置入28天(n=24)及90天(n=12),PFSES与BMS相比,显著降低管腔丢失(0.69±0.49 mm比1.27±0.36 mm,P=0.041;0.77±0.44 mm比1.33±0.29 mm,P<0.01)。在90天组织形态学分析中,损伤积分相似的情况下,PFSES较BMS明显减少新生内膜面积(2.412±1.149 mm2比4.475±1.345 mm2,P<0.05)。PFSES炎症反应与BMS相似,且明显低于PSES。结论PFSES在置入小型猪冠状动脉28及90天后,可以安全有效地抑制内膜增殖和预防支架内再狭窄。
关键词(KeyWords): 支架;西罗莫司;药物载体;冠状动脉再狭窄;猪,雏型
基金项目(Foundation):
作者(Author): 陈明,郑博,彭红玉,王新刚,霍勇
参考文献(References):
- [1]Isenberg MJ.A hierarchical Bayesian meta-analysis of random-ised clinical trials of drug-eluting stents.Lancet,2004,364:583-591.
- [2]Sousa JE,Costa MA,Sousa AG,et al.Two-year angiographicand intravascular ultrasound follow-up after implantation of siroli-mus-eluting stents in human coronary arteries.Circulation,2003,107:381-383.
- [3]Morice MC,Serruys PW,Sousa JE,et al.A randomized com-parison of a sirolimus-eluting stent with a standard stent for coro-nary revascularization.N Engl J Med,2002,346:1773-1780.
- [4]Moses J,Leon MB,Popma JJ,et al.Sirolimus-eluting stentsversus standard stents in patients with stenosis in a native coro-nary artery.N Engl J Med,2003,349:1315-1323.
- [5]Stone GW,Ellis SG,CoxDA,et al.Apolymer-based,paclitax-el-eluting stent in patients with coronary artery disease.N Engl JMed,2004,350:221-231.
- [6]Virmani R,Guagliumi G,Farb A,et al.Localized hypersensitiv-ity and late coronary thrombosis secondary to a sirolimus-elutingstent:Should we be cautious?Circulation,2004,109:701-705.
- [7]Virmani R,Farb A,Guagliumi G,et al.Drug-eluting stents:Caution and concerns for Long-term outcome.Coron Artery Dis,2004,15:313-331.
- [8]McFadden EP,Stabile E,Regar E,et al.Late thrombosis indrug-eluting coronary stents after discontinuation of antiplatelettherapy.Lancet,2004,364:1519-1521.
- [9]Fernandez C,Hernandez R.Drug-eluting stent thrombosis:Re-sults from a pooled analysis including 10 randomized studies.JAm Coll Cardiol,2005,45;954-959.
- [10]Wessely R,Hausleiter J,Michaelis C,et al.Inhibition of neoin-tima for mation by a novel drug-eluting stent systemthat allows fordoseadjustable,multiple,and on-site stent coating.ArteriosclerThromb Vasc Biol,2005,25:748-753.
- [11]Schwartz RS,Huber KC,Murphy JG,et al.Restenosis and theproportional neointimal response to coronary artery injury:Resultsin a porcine model.J Am Coll Cardiol,1992,19:267-274.
- [12]Cilingiroglu M,Elliott J,Patel D,et al.Long-term effects of no-vel biolimus eluting DEVAX AXXESS plus nitinol self-expandingstent in a porcine coronary model.Cathet Cardiovasc Interv,2006,68:271-279.
- [13]Carter AJ,Aggarwal M,Kopia GA,et al.Long-term effects ofpolymer-based,slow-release,sirolimus-eluting stents in a porcinecoronary model.Cardiovasc Res,2004,63:617-624.
- [14]Virmani R,Kolodgie FD,Farb A.Drug-eluting stents:are theyreally safe?Am Heart Hosp J,2004,2:85-88.