不同对比剂浓度对准分子激光消蚀作用的影响:体外研究The impact of different contrast agent concentrations on excimer laser ablation: an in vitro study
何攀,史杨,陈海威,杨俊杰,郭军,陈韵岱,汪奇
摘要(Abstract):
目的 探讨对比剂浓度对准分子激光消蚀斑块作用的影响。方法 采用直径为0.9mm的激光纤维导管,运用波长为308nm的氯化氙准分子激光进行斑块模型消蚀实验。在准分子激光消蚀过程中,根据所注射对比剂浓度不同分为5组:生理盐水组、25%浓度组、50%浓度组、75%浓度组、100%浓度组。通过光学相干断层成像评估准分子激光消蚀后斑块管腔面积变化,以评估对比剂浓度对准分子激光消蚀作用的影响。同时,利用水听器测量准分子激光在不同对比剂浓度的液体中产生的冲击波压力,以探讨准分子激光的冲击波压力与其消蚀作用的相关性。结果 75%浓度组和100%浓度组的消蚀面积相近(P>0.05),均大于50%浓度组、25%浓度组和生理盐水组(均P <0.001)。50%浓度组的消蚀面积显著大于25%浓度组和生理盐水组(均P<0.001),而25%浓度组则大于生理盐水组(P<0.001)。对比剂浓度对准分子激光冲击波压力的影响也呈现相似的趋势。准分子激光产生的冲击波压力与其消蚀面积之间存在显著正相关(r=0.9987,P<0.001)。结论 可以通过改变对比剂浓度来调节准分子激光对斑块组织消蚀作用的强度,为准分子激光联合对比剂注射技术在冠心病治疗中运用提供了指导。
关键词(KeyWords): 准分子激光;对比剂注射;消蚀作用;光学相干断层成像
基金项目(Foundation):
作者(Author): 何攀,史杨,陈海威,杨俊杰,郭军,陈韵岱,汪奇
参考文献(References):
- [1]Tsutsui RS,Sammour Y,Kalra A,et al.Excimer laser atherectomy in percutaneous coronary intervention:a contemporary review[J].Cardiovasc Revasc Med,2021,25:75-85.DOI:10.1016/j.carrev.2020.10.016.
- [2]Protty MB,Hussain HI,Gallagher S,et al.Excimer laser coronary atherectomy during complex PCI:an analysis of1,471 laser cases from the British Cardiovascular Intervention Society database[J].Catheter Cardiovasc Interv,2021,97(5):E653-E660.DOI:10.1002/ccd.29251.
- [3]Karacsonyi J,Alaswad K,Choi JW,et al.Laser for balloon uncrossable and undilatable chronic total occlusion intervrentions[J].Int J Cardiol,2021,336:33-37.DOI:10.1016/j.ijcard.2021.05.015.
- [4]王宁远,冯欢欢,王威,等.光学相干断层成像指导准分子激光联合药物涂层球囊治疗复发性支架内再狭窄的初步研究[J].中国介入心脏病学杂志,2023,31(8):579-584.DOI:10.3969/j.issn.1004-8812.2023.08.003.
- [5]何攀,汪奇.准分子激光辅助治疗冠状动脉支架内再狭窄的研究进展[J].中国介入心脏病学杂志,2022,30(9):693-696.DOI:10.3969/j.issn.1004-8812.2022.09.010.
- [6]付博,赵小丽,张晗,等.激光技术在血栓消融中的应用与进展[J].中国激光,2022,49(19):207-218.DOI:10.3788/cjl202249.1907001.
- [7]Mastuda J.Adjunctive effect of excimer laser coronary angioplastyfor treatment of in-stent restenosis with durg-coated ballon[J].J Am Coll Cardiol,2018,71(11):A1374.DOI:10.1016/S0735-1097(18)31915-6.
- [8]Miyazaki T,Ashikaga T,Fukushima T,et al.Treatment of in-stent restenosis by excimer laser coronary atherectomy and drug-coated balloon:serial assessment with optical coherence tomography[J].J Interv Cardiol,2019,2019:6515129.DOI:10.1155/2019/6515129.
- [9]Ashikaga T,Yoshikawa S,Isobe M.The effectiveness of excimer laser coronary atherectomy with contrast medium for underexpanded stent:the findings of optical frequency domain imaging[J].Catheter Cardiovasc Interv,2015,86(5):946-949.DOI:10.1002/ccd.25915.
- [10]Golino L,Caiazzo G,CalabróP,et al.Excimer laser technology in percutaneous coronary interventions:cardiovascular laser society's position paper[J].Int J Cardiol,2022,350:19-26.DOI:10.1016/j.ijcard.2021.12.054.
- [11]Baumbach A,Haase KK,Rose C,et al.Formation of pressure waves during in vitro excimer laser irradiation in whole blood and the effect of dilution with contrast media and saline[J].Lasers Surg Med,1994,14(1):3-6.DOI:10.1002/lsm.1900140104.
- [12]Herzog A,Steinberg I,Gaisenberg E,et al.A route to laser angioplasty in the presence of fluoroscopy contrast media,using a nanosecond-pulsed 35 5-nm laser[J].IEEE Journal of Selected Topics in Quantum Electronics,2015,22(3):342-347.DOI:10.1109/jstqe.2015.2487889.
- [13]White PF,Elvir Lazo OL,Galeas L,et al.Use of electroanalgesia and laser therapies as alternatives to opioids for acute and chronic pain management[J].F1000Res,2017,6:2161.DOI:10.1268 8/f1000research.12324.1.
- [14]Fales AM,Vogt WC,Wear KA,et al.Experimental investigation of parameters influencing plasmonic nanoparticlemediated bubble generation with nanosecond laser pulses[J].J Biomed Opt,2019,24(6):1-10.DOI:10.1117/1.JBO.24.6.065003.
- [15]Lokar z,Horvat D,Petelin J,et al.Ultrafast measurement of laser-induced shock waves[J].Photoacoustics,2023,30:100465.DOI:10.1016/j.pacs.2023.100465.
- [16]Deckelbaum LI,Natarajan MK,Bittl JA,et al.Effect of intracoronary saline infusion on dissection during excimer laser coronary angioplasty:a randomized trial.The Percutaneous Excimer Laser Coronary Angioplasty(PELCA)Investigators[J].J Am Coll Cardiol,1995,26(5):1264-1269.DOI:10.1016/0735-1097(95)00330-4.
- [17]Sato T,Tsuchida K,Yuasa S,et al.The effect of the debulking by excimer laser coronary angioplasty on long-term outcome compared with drug-coating balloon:insights from optical frequency domain imaging analysis[J].Lasers Med Sci,2020,35(2):403-412.DOI:10.1007/s10103-019-02833-1.
- [18]Lee T,Shlofmitz RA,Song L,et al.The effectiveness of excimer laser angioplasty to treat coronary in-stent resteno sis with peri-stent calcium as assessed by optical coherence tomography[J].EuroIntervention,2019,15(3):e279-e288.DOI:10.4244/EIJ-D-18-00139.
- [19]Mohandes M,Pernigotti A,Moreno C,et al.Coronary laser with simultaneous contrast injection for the treatment of stent underexpansion[J].Cardiol J,2023.DOI:10.5603/cj.96438.