左室主动固定电极在心脏再同步治疗中的初步应用Primary experience with an active fixation coronary sinus lead for cardiac resynchronization therapy
周菁,丁燕生,李康,盛琴慧,蒋捷,王禹川,史力斌
摘要(Abstract):
目的初步探讨左室主动固定电极在心脏再同步治疗(CRT)中的应用价值。方法选取5例符合心脏再同步治疗适应证的心力衰竭(心衰)患者,在常规CRT植入技术基础上,左室电极选用美敦力主动固定电极Attain Star Fix4195,到位后伸展伞叶将其固定至目标血管位置,测试参数并检测有无膈肌刺激,随访观察有无膈肌刺激、脱位、阈值明显增高及心功能改善情况。结果 5例心衰患者全部手术成功。靶目标血管分别为心侧静脉2例,心后侧静脉1例,心中静脉2例。起搏阈值为(1.3±0.7)V,R波幅度(7.9±2.9)mV。术后1周超声心动图与术前比较:左室舒张末内径(LVDd)由(7.43±0.87)cm下降为(7.15±0.76)cm,左室射血分数(LVEF)由(28.8±2.8)%上升为(36.0±4.0)%,肺动脉压(PAPs)由(51.0±20.6)mm Hg下降为(24.6±2.6)mm Hg,二尖瓣反流均有不同程度减轻。术后6~16个月随访:NYHA分级4例患者由Ⅲ级下降为Ⅱ级,1例患者由Ⅳ级下降为Ⅲ级,均无脱位和膈肌刺激出现。结论 CRT采用左室主动固定电极安全性好,降低脱位和膈肌刺激风险,增加手术成功率。
关键词(KeyWords): 电生理学技术,心脏心力衰竭;电极,植入
基金项目(Foundation):
作者(Author): 周菁,丁燕生,李康,盛琴慧,蒋捷,王禹川,史力斌
参考文献(References):
- [1]Jarcho JA.Biventricular pacing.N Engl J Med,2006,355:288-294.
- [2]Borleffs CJ,van Bommel RJ,Molhoek SG,et al.Requirement for coronary sinus lead interventions and effectiveness of endovascular replacement during long-term follow-up after implantation of a resynchronization device.Europace,2009,11:544-545.
- [3]León AR,Abraham WT,Curtis AB,et al.MIRACLE Study Program.Safety of transvenous cardiac resynchronization system implantation in patients with chronic heart failure:combined results of over2,000patients from a multicenter study program.J Am Coll Cardiol,2005,46:2348-2356.
- [4]臧红云,王冬梅,韩雅玲,等.110例心脏再同步化治疗中左室电极导线植入的临床经验和并发症分析.中国心脏起博与心电生理杂志,2009,23:480-483.
- [5]Ngele H,Azizi M,Hashagen S,et al.First experience with a new active fixation coronary sinus lead.Europace,2007,9:437-441.
- [6]De Cock CC,Jessurun ER,Allaart CA,et al.Repetitive intraoperative dislocation during transvenous lead implantation:usefulness of the retained guidewire technique.Pacing Clin Electrophysiol,2004,27:1589-1593.
- [7]Ngele H,Hashagen S,Ergin M,et al.Coronary sinus lead fragmentation2years after implantation with a retained guidewire.Pacing Clin Electrophysiol,2007,30:438-439.
- [8]Kowalski O,Prokopczuk J,Lenarczyk R,et al.Coronary sinus stenting for the stabilization of left ventricular lead during resynchronization therapy.Europace,2006,8:367-370.
- [9]Luedorff G,Kranig W,Grove R,et al.Improved success rate of cardiac resynchronization therapy implant by employing an active fixation coronary sinus lead.Europace,2010,12:825-829.
- [10]Crossley GH,Exner D,Mead RH,et al.Chronic performance of an active fixation coronary sinus lead.Heart Rhythm,2010,7:472-478.